Customed, Inc: Medical Device Recall in 2016 - (Recall #: Z-1427-2016)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2016.
Data Source: FDA.
Product Description:
Catalog Number: 9001496 D&C LAPAROSCOPY SURGICAL PACK Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc.
Product Classification:
Class II
Date Initiated: January 7, 2016
Date Posted: April 20, 2016
Recall Number: Z-1427-2016
Event ID: 72992
Reason for Recall:
Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.
Status: Terminated
Product Quantity: 12
Code Information:
LOTS: SL15181 (150619167) 30-JUN-16 SL15181 (150619386) 30-JUN-16 SL15209 (150619386) 30-JUN-16 SL15209 (150719646) 31-JUL-16 SL15224 (150820077) 31-AUG-16
Distribution Pattern:
Distributed only in Puerto Rico.
Voluntary or Mandated:
Voluntary: Firm initiated