Customed, Inc: Medical Device Recall in 2016 - (Recall #: Z-1487-2016)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2016.
Data Source: FDA.
Product Description:
Catalog Number: 900774 INCISION AND DRAINAGE TRAY Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc.
Product Classification:
Class II
Date Initiated: January 7, 2016
Date Posted: April 20, 2016
Recall Number: Z-1487-2016
Event ID: 72992
Reason for Recall:
Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.
Status: Terminated
Product Quantity: 800
Code Information:
LOTS: SL15177 (150619315) 30-JUN-16 SL15268 (150719854) 30-SEP-16 SL15271 (150719854) 30-SEP-16
Distribution Pattern:
Distributed only in Puerto Rico.
Voluntary or Mandated:
Voluntary: Firm initiated