Cytocell Ltd.: Medical Device Recall in 2022 - (Recall #: Z-0824-2022)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2022.
Data Source: FDA.
Product Description:
Cytocell 8 Square Template Slides (Glass)- microscope slides Reference: PCN008
Product Classification:
Class III
Date Initiated: February 14, 2022
Date Posted: March 30, 2022
Recall Number: Z-0824-2022
Event ID: 89747
Reason for Recall:
Labelled with an incorrect expiry date on the outer label with an extended expiry date than the validated shelf life. The correct expiry date for the device is 2022-02 and not 2023-05 as shown on the outer label.
Status: Terminated
Product Quantity: 5 kits
Code Information:
Lot 210511-018
Distribution Pattern:
US Nationwide distribution in the states of CT, TX.
Voluntary or Mandated:
Voluntary: Firm initiated