Cytrellis Biosystems, Inc.: Medical Device Recall in 2023 - (Recall #: Z-1648-2023)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2023.
Data Source: FDA.
Product Description:
Cytrellis ellacor System with Micro-Coring Technology-Dermal Micro-Coring System System intended for the treatment of moderate and severe wrinkles in the mid and lower face. Model #: Gen 4
Product Classification:
Class II
Date Initiated: April 7, 2023
Date Posted: June 7, 2023
Recall Number: Z-1648-2023
Event ID: 92187
Reason for Recall:
Irregular core patterns could cause potential overlap of cores at the 8% setting, potentially leading to prolonged healing, irregularities in patient skin texture post-treatment, and/or scarring.
Status: Completed
Product Quantity: 89 units
Code Information:
UDI-DI: B933ELCON010 All serial numbers
Distribution Pattern:
Nationwide
Voluntary or Mandated:
Voluntary: Firm initiated