Dako Denmark A/S: Medical Device Recall in 2016 - (Recall #: Z-2330-2016)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2016.
Data Source: FDA.
Product Description:
EnVision FLEX+, Mouse, High pH (Link), Product No. K8002, K8004. Intended for use in immunohistochemistry together with Autostainer Link Instruments.
Product Classification:
Class II
Date Initiated: July 6, 2016
Date Posted: August 10, 2016
Recall Number: Z-2330-2016
Event ID: 74608
Reason for Recall:
The pH of Target Retrieval Solution in the kit measures 4-5 instead of the correct pH of 9.1.
Status: Terminated
Product Quantity: 147 kits
Code Information:
K8002 Lot no: 10112879 K8004 Lot no: 10111463
Distribution Pattern:
Worldwide Distribution -- US, Canada, and Colombia.
Voluntary or Mandated:
Voluntary: Firm initiated