Dako North America Inc.: Medical Device Recall in 2015 - (Recall #: Z-1261-2015)
See the recall detail below. You can also see other recalls from the same firm in 2015.
Autostainer Link 48 (AS480), Autostainer Plus Link Instrument (AS100), Autostainer (S3400), and Autostainer Plus (S3800). Automated slide stainer for in vitro diagnostic use.
Class II
A false negative result affecting the diagnosis may occur. A defect in the syringe assembly stopcock component may, in some circumstances, cause leakage of excess buffer onto slide location 34 on the AS480, or slide locations 35 and 36 on the AS100, S3400, and S3800. Affected dates are from 12/2013-08/2014.
992640 Rev. 02, 992085 Rev.no H, 992625 Rev.no F, 992326 Rev.no C Affected dates: 12/2013-08/2014 Each device is labeled with a unique serial number. Catalog/Model Number: - AS480 - S3800 - S3400 - AS100
Worldwide Distribution: US (nationwide) and countries of: AT, AU, BE, CA, CH, DE, DK, ES, FR, GB, IE, IT, JP, NL, NO, PL, and SE.
Voluntary: Firm initiated