Data Innovations, Inc.: Medical Device Recall in 2014 - (Recall #: Z-0571-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

Instrument Manager driver : Driver sysu100i that connects to the Sysmex UF-100 analyzer Software all versions prior to( v7.00.0006 and v8.00.0005): v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004

Product Classification:

Class II

Date Initiated: November 6, 2013
Date Posted: January 1, 2014
Recall Number: Z-0571-2014
Event ID: 66754
Reason for Recall:

When connected to the Sysmex UF-100, the Instrument Manager driver sysu100i driver software versions 7.00.0001, 7.00.0002, 7.00.0003, 7.00.0004, 7.00.0005, 8.00.0001, 8.00.0002, 8.00.0003, and 8.00.0004, and is configured with option "Strip Zeros from Result" set to "Yes" may result in sample error

Status: Terminated
Product Quantity: 24 systems
Code Information:

Software alll versions prior to: v7.00.0006 and v8.00. Includes: v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004

Distribution Pattern:

Nationwide Canada and Britain

Voluntary or Mandated:

Voluntary: Firm initiated