Datascope Corp.: Medical Device Recall in 2023 - (Recall #: Z-1145-2023)
See the recall detail below. You can also see other recalls from the same firm in 2023.
Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Containing Coiled Cord cable (part number 0012-00-1801, Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-UC-0800-31, 0998-UC-0800-33, 0998-UC-0800-52, 0998-UC-0800-53, 0998-00-0800-55
Class I
An unexpected shutdown of the IABP may occur due to a failure of the connection between the Coiled Cord cable and the Cable Assembly backplane to the Coiled Cord cable which provides the communication between the display head and base unit.
All serial numbers. Model 0998-00-0800-31, UDI-DI 10607567109053; Model 0998-00-0800-32, UDI-DI 10607567111117; Model 0998-00-0800-33, UDI-DI 10607567109008; Model 0998-00-0800-34, UDI-DI 10607567111940; Model 0998-00-0800-35, UDI-DI 10607567109107; Model 0998-00-0800-45, UDI-DI 10607567108421; Model 0998-00-0800-52, UDI-DI 10607567108438; Model 0998-00-0800-53, UDI-DI 10607567108391; Model 0998-00-0800-55, UDI-DI 10607567108414; Model 0998-00-0800-65, UDI-DI 10607567113432
Worldwide Distribution: US (nationwide) and OUS (Foreign) ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BRAZIL, BRUNEI , BULGARIA, CANADA, CHILE, CHINA, CZECH REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, GHANA, HONG KONG, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KUWAIT, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAYSIA, MEXICO, MOROCCO, MYANMAR, NETHERLANDS, NIGERIA, NORWAY, OMAN, PAKISTAN, PALESTINE, PHILIPPINES, POLAND, PORTUGAL, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, THAILAND, TRINIDAD AND TOBAGO, TURKEY, TURKMENISTAN, UNITED ARAB EMIRATES, UNITED KINGDOM, VIETNAM, AND YEMEN.
Voluntary: Firm initiated