Deerfield Imaging, Inc.: Medical Device Recall in 2018 - (Recall #: Z-0975-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

IMRIS ORT100 Table, Part Number 110470-000

Product Classification:

Class II

Date Initiated: February 20, 2018
Date Posted: March 21, 2018
Recall Number: Z-0975-2018
Event ID: 79360
Reason for Recall:

The rotational locking mechanisms on the ORT 100/ T2X tables can become disengaged or unlocked while in use which could potentially result in unintended movement of the table.

Status: Terminated
Product Quantity: 15 tables
Code Information:

Serial Numbers: 10002341, 10003629, 10003801, 10002654, 10002795, 10002796, 10002847, 10003830, 10003243, 10002848, 10003241, 10005651, 10007348, 10007424, 10006886

Distribution Pattern:

US, Canada, Australia, China, Qatar

Voluntary or Mandated:

Voluntary: Firm initiated