Deerfield Imaging, Inc.: Medical Device Recall in 2020 - (Recall #: Z-0762-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

iMRX MRI and Angiography System, part number 10094137 and 10094141, a surgical/interventional suite with integrated MRI and X-ray angiography. The IMRIS iMRX system is the integration of a Siemens Artis zee family (biplane or floor or ceiling mounted single plane) and the IMRIS Neuro 3T System. The System is a solution providing MRI scanning capabilities in the angiography room while retaining all functionality of Siemens Artis zee system. The System includes a traditional MRI unit that has been suspended on an overhead rail system, and is designed to operate inside RF room to facilitate intra-operative, interventional, and multi-room imaging. It also allows intra-operative or interventional imaging capabilities, while retaining all diagnostic functionality of a standard MRI system in the diagnostic room. The iMRX system allows for the sequential use of either MR imaging or X-ray fluoroscopic imaging of the head or chest for a single patient while positioned on a common fixed table either prior to, during or post clinical procedure. When these images are interpreted by a trained physician, they may assist in diagnosis during either neurovascular or cardiovascular interventionsa procedures. The iMRX system is designed such that the imaging sub-systems cannot operate simultaneously and only in accordance with their respective cleared intended use.

Product Classification:

Class II

Date Initiated: November 26, 2019
Date Posted: January 8, 2020
Recall Number: Z-0762-2020
Event ID: 84400
Reason for Recall:

There is a potential hardware error for the iMRX System with a generator A100. A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible

Status: Terminated
Product Quantity: 6
Code Information:

Serial Number: 146792, 153374, 153642, 153793, 153896, 154588.

Distribution Pattern:

CT, MA, VA. International distribution: Australia, Canada, and Qatar.

Voluntary or Mandated:

Voluntary: Firm initiated