Deerfield Imaging, Inc.: Medical Device Recall in 2021 - (Recall #: Z-2586-2021)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2021.
Data Source: FDA.
Product Description:
Operating room table - Product Usage: intended for use during diagnostic examinations, radiologic procedures or surgical procedures to support and position a patient.
Product Classification:
Class II
Date Initiated: August 25, 2021
Date Posted: October 6, 2021
Recall Number: Z-2586-2021
Event ID: 88661
Reason for Recall:
ORT200/300 rotational lock may fail to function such that it cannot change states between locked and unlocked, due to a mechanical hardware issue.
Status: Terminated
Product Quantity: 3 units
Code Information:
UDI: 00857534006462; Model Number: 114093-600; Serial Number: 10007411, 10002996, 10004515.
Distribution Pattern:
Worldwide distribution - US Nationwide distribution FL, NH, TX, OH, CA, OR, UT, TN, NC and the countries of Canada, Australia, Sweden, Japan.
Voluntary or Mandated:
Voluntary: Firm initiated