Degania Medical Devices Pvt. Ltd.: Medical Device Recall in 2020 - (Recall #: Z-1782-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

Temperature Sensing 100% Silicone Foley Catheter, Part Number 102201101463MD - Product Usage: is a 3 lumen silicone catheter with a sensor to measure temperature while draining the urinary bladder.

Product Classification:

Class II

Date Initiated: March 17, 2020
Date Posted: April 29, 2020
Recall Number: Z-1782-2020
Event ID: 85327
Reason for Recall:

Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the specified lots of temperature sensing catheters.

Status: Terminated
Product Quantity: 16393
Code Information:

Lot Numbers: V18022033 V18025110 V18025111 V18029225 V18029226 V19024273 V19013327 V19025875 V19016720 V19025874

Distribution Pattern:

Worldwide distribution - US Nationwide distribution including state of IL and the country of Mexico.

Voluntary or Mandated:

Voluntary: Firm initiated