Degania Medical Devices Pvt. Ltd.: Medical Device Recall in 2020 - (Recall #: Z-1785-2020)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2020.
Data Source: FDA.
Product Description:
Temperature Sensing 100% Silicone Foley Catheter, Part Number 102202101663DO - Product Usage: is a 3 lumen silicone catheter with a sensor to measure temperature while draining the urinary bladder.
Product Classification:
Class II
Date Initiated: March 17, 2020
Date Posted: April 29, 2020
Recall Number: Z-1785-2020
Event ID: 85327
Reason for Recall:
Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the specified lots of temperature sensing catheters.
Status: Terminated
Product Quantity: 34785
Code Information:
Lot Numbers: V18013621 V18014635 V18025989 V18029252 V18032217 V18031184 V19006387 V18049356 V18049355 V19008749
Distribution Pattern:
Worldwide distribution - US Nationwide distribution including state of IL and the country of Mexico.
Voluntary or Mandated:
Voluntary: Firm initiated