Degania Silicone, Ltd.: Medical Device Recall in 2018 - (Recall #: Z-2783-2018)
See the recall detail below. You can also see other recalls from the same firm in 2018.
Temperature sensing catheter 18FR, Catalogue Number 102203101863DO Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
Class II
The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
V17015592, V17016065
Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
Voluntary: Firm initiated