DePuy Mitek, Inc., a Johnson & Johnson Co.: Medical Device Recall in 2017 - (Recall #: Z-2617-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2017.

Data Source: FDA.
Product Description:

DePuy Mitek LATARJET EXPERIENCE Coracoid Top Hat Drill Product Code: 288202 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by supporting the anteroinferior glenoid with a bony graft

Product Classification:

Class II

Date Initiated: May 12, 2017
Date Posted: June 28, 2017
Recall Number: Z-2617-2017
Event ID: 77252
Reason for Recall:

Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw

Status: Terminated
Product Quantity: 184 units
Code Information:

GTN: 01)10886705026807 Lot codes: 16C01, 16E03, 16R01, 17A01, 17B01

Distribution Pattern:

US: AZ, CA, CO, GA, MA, MI, OH, WA Foreign: Austria, Australia, Belgium, France, Germany, Spain, Switzerland, Netherlands, Poland, UK

Voluntary or Mandated:

Voluntary: Firm initiated