DePuy Orthopaedics, Inc.: Medical Device Recall in 2014 - (Recall #: Z-0199-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

ATTUNE Intuition Distal Femoral Jig; sold non-sterile; product is wrapped in bubble wrap and then sealed inside a polyethylene bag

Product Classification:

Class II

Date Initiated: October 15, 2014
Date Posted: November 19, 2014
Recall Number: Z-0199-2015
Event ID: 69271
Reason for Recall:

Issuing a device correction because if the pin bushing is over loaded or inverted, there is the potential for partial or full displacement and fracture of the plastic around the pin bushing.

Status: Terminated
Product Quantity: 3,082
Code Information:

Catalog NOs. 2544-00-521 and 2544-00-520; All lots are affected for the Device Correction. Lot numbers below should be inspected and swapped out if the bushing is incorrectly assembled: Part No. DePuy Synthes Lot No: MFG Lot No: 254400520 ABC22844 C22844 ABC22845 C22845 ABC34279 C34279 ABC34280 C34280 ABC4107 C4107 ABC4108 C4108 ABC4112 C4112 ABC4113 C4113 ABC50741 C50471 254400521 ABB13662 B13662 ABB19116 B19116 ABB19117 B19117 ABB19118 B19118 ABB35392 B35392 ABB35393 B35393 ABB39835 B39835 ABB41384 B41384 ABB46872 B46872 ABB49200 B49200 ABB50724 B50724 ABB55749 B55749 ABB63598 B63598 ABB66346 B66346 ABB72107 B72107 ABB77183 B77183 ABB79390 B79390 ABB79391 B79391 ABB80712 N80712 ABC17165 C17165 ABC5023 C5023

Distribution Pattern:

Worldwide Distribution - USA (nationwide) except AK, including Puerto Rico and the countries of Canada, Australia, Austria, Belgium, China, France, Germany, India, Ireland, Israel, Italy, Japan, Malaysia, Netherlands, New Zealand, Singapore, South Africa, South Korea, Spain, Switzerland, Thailand, UAE, United Kingdom and Chile.

Voluntary or Mandated:

Voluntary: Firm initiated