DePuy Orthopaedics, Inc.: Medical Device Recall in 2022 - (Recall #: Z-0490-2022)
See the recall detail below. You can also see other recalls from the same firm in 2022.
GLOBAL UNITE STD STEM SZ 16 Product Code: 110016100
Class II
Stems may have an undersized spigot bore diameter therefore separating the proximal body from the stem may not be possible without surgical delay and/or bone damage
Lot #: 9793946, 9793947, 9802315, 9802316, 9802317, 9802318, 9802320, 9813856, 9814516, 9824502, 9824503, 9824505, 9834401, 9834402, 9836078, 9841198, 9880972, 9886688, 9887451, 9887452 GTIN: 10603295004189
Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Denmark, Estonia, Finland, France, Germany, Greece, Ireland, Israel, Italy, Netherlands, New Zealand, Norway, Poland, South Africa, Sweden, Switzerland, Thailand, United Kingdom.
Voluntary: Firm initiated