DePuy Orthopaedics, Inc.: Medical Device Recall in 2022 - (Recall #: Z-1419-2022)
See the recall detail below. You can also see other recalls from the same firm in 2022.
ATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in preparation for the femoral implant Part Number: 254400525
Class II
Manufactured with a Galvanized High Carbon Steel spring instead of the correct 302 Stainless steel spring. has the potential to oxidize and corrode in the ATTUNE Measured Sizing & Rotation Guide, resulting in surgical delay, debris in surgical field, Osteolysis, Poor Joint Mechanics, Soft Tissue Irritation, Pain
GTIN: 10603295434061 Lots: AB4536235, AB4733296, AB4841493, AB4810808, AB4841495, AB4904565, AB4967214, AB5004557, and AB5034303
Nationwide Foreign: Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Lithuania, Mauritius, Mexico, Morocco, Netherlands, New Zealand, Norway, Peru, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Vietnam
Voluntary: Firm initiated