DePuy Spine, Inc.: Medical Device Recall in 2013 - (Recall #: Z-0685-2013)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2013.
Data Source: FDA.
Product Description:
DePuy Spine EXPEDIUM Offset Torque Wrench Product Code 2797-29-510 The device is intended for use during the final tightening of an Expedium Offset Spinal Construct.
Product Classification:
Class II
Date Initiated: July 24, 2012
Date Posted: January 23, 2013
Recall Number: Z-0685-2013
Event ID: 62719
Reason for Recall:
A component within the handle of the Expedium Offset Torque Wrench is prematurely wearing, potentially resulting in the instrument recording torque values outside of the specification.
Status: Terminated
Product Quantity: 140 units
Code Information:
All lots
Distribution Pattern:
Worldwide Distribution - USA (nationwide) and internationally to Belgium, Egypt, Ireland, Italy, Portugal, and the UK.
Voluntary or Mandated:
Voluntary: Firm initiated