DePuy Spine, Inc.: Medical Device Recall in 2019 - (Recall #: Z-1225-2019)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2019.
Data Source: FDA.
Product Description:
Concorde Lift, Expandable Interbody Device, spinal fusion implant Convex 9x21 (197809021C), 9x26 (197809026C), 11x21 (197811021C), 11x26 (197811026C), 9x21 (US197809021C), 9x26 (US197809026C)
Product Classification:
Class II
Date Initiated: April 8, 2019
Date Posted: May 8, 2019
Recall Number: Z-1225-2019
Event ID: 82587
Reason for Recall:
A complaint trend was observed for post-operative loss of height and migration for Concorde Lift G3+ implants.
Status: Terminated
Product Quantity: 3166 total
Code Information:
All lots
Distribution Pattern:
US Nationwide distribution. Foreign distribution to Austria, Switzerland, and United Kingdom.
Voluntary or Mandated:
Voluntary: Firm initiated