DeRoyal Industries Inc: Medical Device Recall in 2012 - (Recall #: Z-2008-2012)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2012.

Data Source: FDA.
Product Description:

Custom surgical kits and trays: GEO-MED TOTAL KNEE PACK PGYBK, REF 89-6664.04 GEO-MED TOTAL HIP PACK, REF 89-6664.03 surgical kits

Product Classification:

Class II

Date Initiated: June 16, 2012
Date Posted: July 25, 2012
Recall Number: Z-2008-2012
Event ID: 62457
Reason for Recall:

Surgical kits and trays contain Stryker Hytrel Togas which were recalled because the clear tape, that is intended to aid in the prevention of patient's fluids from contacting the surgeon/surgical staff through the sleeve seam of the toga, was not applied during the manufacturing process.

Status: Terminated
Product Quantity: 303 units
Code Information:

GEO-MED TOTAL KNEE PACK PGYBK, REF 89-6664.04; Lot numbers: 22737521, 22924079, 23427293, 23536123, 24076552, 24435084, 24435084, 24750543, 25088133, 25275511, 25598563, 26476576, 26924324, 26959148, 27264233, 27526176, 27756781, 27756781, 28238361, 28481694 GEO-MED TOTAL HIP PACK, REF 89-6664.03; Lot numbers: 22924204, 23623163, 23880637, 24076691, 24162522, 24295421, 24585728, 24600836, 24900118, 25030484, 25144351, 25461743, 25559311, 26021993, 26064301, 27526328, 28262740

Distribution Pattern:

Nationwide Distribution including SD and NE.

Voluntary or Mandated:

Voluntary: Firm initiated