DeRoyal Industries Inc: Medical Device Recall in 2014 - (Recall #: Z-0834-2014)
See the recall detail below. You can also see other recalls from the same firm in 2014.
Aquarius (TM) TEMPERATURE THERAPY, REF T650, Rx Only, Manufactured for DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, Made in China. Physical Therapy.
Class II
The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.
Lot Numbers: 31480713, 31480721, 31968743, 31971619, 31990844, 32005711, 32018491, 32037764, 32118020, 32710832, 32736223, 33053264, 33105221, 33141176, 33164204, 33164407, 33276273, 33329911
Nationwide Distribution.
Voluntary: Firm initiated