DeRoyal Industries Inc: Medical Device Recall in 2014 - (Recall #: Z-0835-2014)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2014.
Data Source: FDA.
Product Description:
DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee/Shoulder Blanket NS, REF T652NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy.
Product Classification:
Class II
Date Initiated: September 17, 2013
Date Posted: February 5, 2014
Recall Number: Z-0835-2014
Event ID: 66742
Reason for Recall:
The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.
Status: Terminated
Product Quantity: 121 units
Code Information:
Lot Numbers: 32871821, 32976471, 33105247
Distribution Pattern:
Nationwide Distribution.
Voluntary or Mandated:
Voluntary: Firm initiated