DeRoyal Industries Inc: Medical Device Recall in 2023 - (Recall #: Z-1283-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

DeRoyal SafeLiner Suction Canister, 1000cc, REF PHESL-1000B

Product Classification:

Class II

Date Initiated: February 9, 2023
Date Posted: March 29, 2023
Recall Number: Z-1283-2023
Event ID: 91760
Reason for Recall:

The reason for the voluntary recall is due to shrinkage of the canister lid. This shrinkage causes the lid not to contact the outer shell to create a seal and thus, it does not create and hold vacuum.

Status: Ongoing
Product Quantity: 27,200 devices
Code Information:

GTIN 50749756377046, Lot 58447585 exp. 12/22/2027

Distribution Pattern:

US Nationwide distribution in the states of Arizona, Texas, Virginia, Pennsylvania, California, Missouri, North Carolina, Illinois, Indiana, Ohio, Arkansas.

Voluntary or Mandated:

Voluntary: Firm initiated