Dexcom Inc: Medical Device Recall in 2020 - (Recall #: Z-1560-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

Dexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracking patterns in persons (age 18 and older) with diabetes.

Product Classification:

Class II

Date Initiated: February 3, 2020
Date Posted: April 1, 2020
Recall Number: Z-1560-2020
Event ID: 84891
Reason for Recall:

The company identified potential interference from hydroxyurea. Patient use of the anti-neoplastic drug may falsely elevate glucose readings on the Continuous Glucose Monitoring Systems.

Status: Terminated
Product Quantity: 3091 Systems
Code Information:

Device Listing Number: D189069 Model: STS-GL-041 Description: G4/G5 Sensor Kit 4 Pack GUDID Registered GTIN: 30386270000010 Lot Number: All Model: STS-RX-041 Description: G4/G5 Sensor Kit 4 Pack Retail GUDID Registered GTIN: 20386270000013 Lot Number: All Model: STS-MC-004 Description: G4/G5 Sensor Kit 4 Pack Medicare GUDID Registered GTIN: 00386270000637 Lot Number: All

Distribution Pattern:

World-wide Distribution

Voluntary or Mandated:

Voluntary: Firm initiated