Product Description:

Brand Name: Dexcom ONE+ Continuous Glucose Monitoring System Product Name: Dexcom ONE+ Android CGM App Model/Catalog Number: SW14245 Software Version: versions 1.4.0 and earlier Product Description: Dexcom ONE+ Androind CGM App Component: Dexcom ONE+ Continuous Glucose Monitoring System

Product Classification:

Class I

Date Initiated: July 24, 2025
Date Posted: September 10, 2025
Recall Number: Z-2449-2025
Event ID: 97262
Reason for Recall:

The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier), henceforth referred to as the App. The affected App does not provide an expected "Sensor Failed" alert when the transmitter sends a "transmitterFailed" error message to the App when the CGM experiences a hardware or firmware failure. Instead, the App ends the CGM sensor session, stops reporting glucose values and displays the Start Sensor screen or No active sensor message without alerting the user, which can lead to missed detection of a hyperglycemic or hypoglycemic event and a delay in treatment.

Status: Ongoing
Product Quantity: 41629
Code Information:

Lot Code: UDI: SW14245 - 00386270004314

Distribution Pattern:

SW12299 G7 Android CGM App Worldwide Distribution - US Nationwide and the countries of Andorra, United Arab Emirates, Austria, Australia, Belgium, Bahrain, Canada, Switzerland, Czech Republic, Germany, Denmark, Estonia, Spain, Finland, France, United Kingdom, Hong Kong, Ireland, Israel, Iceland, Italy, Jordan, Japan, South Korea, Kuwait, Lebanon, Liechtenstein, Luxembourg, Malta, Netherlands, Norway, New Zealand, Oman, Poland, Portugal, Qatar, Saudi Arabia, Sweden, Slovenia, Slovakia, Turkey, South Africa SW12300 G7 iOS CGM App Worldwide Distribution - US Nationwide and the countries of United Arab Emirates, Austria, Australia, Belgium, Bahrain, Canada, Switzerland, Czech Republic, Germany, Denmark, Estonia, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, Iceland, Italy, Jordan, Japan, South Korea, Kuwait, Lebanon, Liechtenstein, Luxembourg, Malta, Netherlands, Norway, New Zealand, Oman, Poland, Portugal, Qatar, Saudi Arabia, Sweden, Slovenia, Slovakia, Turkey, South Africa SW14245 Dexcom ONE+ Android Mobile App OUS worldwide distribution to the countries of Andorra, Australia, Belgium, Bulgaria, Estonia, Spain, France, United Kingdom, Greece, Croatia, Ireland, Israel, Italy, Lithuania, Latvia, Netherlands, New Zealand, Poland, Romania, Saudi Arabia, Sweden SW14244 D1G7 iOS Mobile App OUS worldwide distribution to the countries of Andorra, Australia, Belgium, Bulgaria, Estonia, Spain, France, United Kingdom, Greece, Croatia, Ireland, Italy, Lithuania, Latvia, Netherlands, New Zealand, Poland, Romania, Saudi Arabia, Sweden

Voluntary or Mandated:

Voluntary: Firm initiated