Diasorin Inc.: Medical Device Recall in 2016 - (Recall #: Z-0645-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

LIAISON Estradiol II Gen, 310680, 310680-CN The DiaSorin LIAISON Estradiol II Gen assay is a chemiluminescent immunoassay (CLIA) intended for the quantitative determination of estradiol in human serum. Assay results should be used in conjunction with other clinical and laboratory data to assist in the diagnosis and management of conditions involving an excess or deficiency of estradiol. The test has to be performed on the LIAISON Analyzer Family.

Product Classification:

Class II

Date Initiated: July 13, 2016
Date Posted: November 30, 2016
Recall Number: Z-0645-2017
Event ID: 75478
Reason for Recall:

DiaSorin is issuing an urgent medical device correction for the LIAISON¿ Estradiol II Gen assay. Due to the risk of a recently identified cross reactivity, the LIAISON¿ Estradiol II Gena assay should not be used to test patients being treated with fulvestrant (Faslodex).

Status: Terminated
Product Quantity: 10,739
Code Information:

Catalog Number 31 0680, 31 0680-CN

Distribution Pattern:

USA (nationwide) Distribution and Worldwide Distribution including the states of : AR, AZ, CA, FL, ID, NJ, NV, NY and PA.., and to the countries of : Saudi Arabia, Colombia, Mauritius, Iraq, Mongolia, Greece, Albania, Austria, Bosnia, Bosnia Herzegovina, Belarus, Bulgaria, Macedonia, Poland, Serbia, Dominican Rep., Tunisia, Paraguay, Hungary, Jordan, Egypt, Belgium, Caribbean, Germany, Brazil, Mexico, Italy, India, South Korea, Cyprus, Bangladesh, Morocco, Chile, Bahrain, Palestine, Utd.Arab.Emir., Kosovo, Uruguay, Venezuela, Hungary, Australia, Vietnam, Sri Lanka, Turkey, Vietnam, Romania, Mauritius, Ecuador, Lebanon, Nepal, Argentina, Hong Kong, China and Canada.

Voluntary or Mandated:

Voluntary: Firm initiated