Draeger Medical, Inc.: Medical Device Recall in 2018 - (Recall #: Z-1954-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

Carina Ventilator, ventilator. Catalog no. 5704110. Long-term ventilator for treatment in hospitals by a qualified medical personnel.

Product Classification:

Class II

Date Initiated: April 26, 2018
Date Posted: June 6, 2018
Recall Number: Z-1954-2018
Event ID: 80027
Reason for Recall:

The valve assembly material may become corroded due to insufficient cleaning by a supplier. There is a risk of the breathing gas becoming contaminated in the affected devices.

Status: Terminated
Product Quantity: 8 units
Code Information:

Serial nos. ASKM-0065 ASKM-0066 ASKM-0106 ASKM-0134 ASKM-0137 ASKM-0150 ASKM-0151 ASKN-0009

Distribution Pattern:

Distributed to the states of CT, IL, MN, NV, SD, and WA.

Voluntary or Mandated:

Voluntary: Firm initiated