Draeger Medical, Inc.: Medical Device Recall in 2021 - (Recall #: Z-0108-2022)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2021.
Data Source: FDA.
Product Description:
Evita V500 Ventilator with Installed CO2 Measurement Option-Intended for the ventilation of adult,pediatric and neonatal patients. Catalog Number: 8416400
Product Classification:
Class II
Date Initiated: September 10, 2021
Date Posted: October 20, 2021
Recall Number: Z-0108-2022
Event ID: 88635
Reason for Recall:
Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds
Status: Ongoing
Product Quantity: 6,797 units
Code Information:
Software 2.51 and Lower
Distribution Pattern:
US Nationwide
Voluntary or Mandated:
Voluntary: Firm initiated