Draeger Medical, Inc.: Medical Device Recall in 2021 - (Recall #: Z-0109-2022)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2021.
Data Source: FDA.
Product Description:
Babylog VN500 Ventilator with Installed CO2 Measurement Option-intended for the ventilation of neonatal patients from 0.4 kg (0.88 lbs) up to 10 kg (22 lbs) and pediatric patients from 5 kg (11 lbs) up to 20 kg (44 lbs) bodyweight. Catalog Number: 8417400
Product Classification:
Class II
Date Initiated: September 10, 2021
Date Posted: October 20, 2021
Recall Number: Z-0109-2022
Event ID: 88635
Reason for Recall:
Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds
Status: Ongoing
Product Quantity: 2,169 units
Code Information:
Software 2.51 and Lower
Distribution Pattern:
US Nationwide
Voluntary or Mandated:
Voluntary: Firm initiated