Draeger Medical Systems, Inc.: Medical Device Recall in 2014 - (Recall #: Z-0714-2014)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2014.
Data Source: FDA.
Product Description:
Draeger Infinity Acute Care System Monitoring Solution
Product Classification:
Class II
Date Initiated: December 10, 2013
Date Posted: January 22, 2014
Recall Number: Z-0714-2014
Event ID: 67056
Reason for Recall:
After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.
Status: Terminated
Product Quantity: 439
Code Information:
with software versions VG2.0.3 and higher.
Distribution Pattern:
US Distribution including the states of: MA, NH, IN, CA, TX, NY, OK, and IL.
Voluntary or Mandated:
Voluntary: Firm initiated