Draeger Medical Systems, Inc.: Medical Device Recall in 2014 - (Recall #: Z-1440-2014)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2014.
Data Source: FDA.
Product Description:
Draeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833
Product Classification:
Class I
Date Initiated: March 5, 2014
Date Posted: April 16, 2014
Recall Number: Z-1440-2014
Event ID: 67677
Reason for Recall:
The potentiometers (knobs) for setting the ventilation parameters may have contact interrupted and the device discontinues ventilation, gives an audible alarm and displays the error message Poti unplugged.
Status: Terminated
Product Quantity: 3
Code Information:
Oxylog 3000 plus, Part number: 5704833 with serial numbers: ASEK-0003, ASEK-0004, ASEK-0005
Distribution Pattern:
USA Nationwide Distribution in the states of WI and NY.
Voluntary or Mandated:
Voluntary: Firm initiated