Draeger Medical Systems, Inc.: Medical Device Recall in 2023 - (Recall #: Z-1471-2023)
See the recall detail below. You can also see other recalls from the same firm in 2023.
Draeger Infinity CentralStation (ICS), centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment, Software versions VG2.1.3 and lower
Class II
The software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this condition, ECG waveforms with unusually narrow and/or high frequency QRS may be displayed or printed with QRS amplitudes intermittently lower than actual.
UDI-DI: 04049098001878; All Systems Running Software versions: ICS VG2.1.3 and lower
Worldwide - US Nationwide distribution.
Voluntary: Firm initiated