Draeger Medical Systems, Inc.: Medical Device Recall in 2023 - (Recall #: Z-1472-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

Draeger Infinity M300 and M300+, for use with the ICS to monitor ECG and pulse oximetry on ambulatory and non-ambulatory adult and pediatric patients using wireless communication over the Infinity patient monitoring network,

Product Classification:

Class II

Date Initiated: April 5, 2023
Date Posted: May 3, 2023
Recall Number: Z-1472-2023
Event ID: 91938
Reason for Recall:

The software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this condition, ECG waveforms with unusually narrow and/or high frequency QRS may be displayed or printed with QRS amplitudes intermittently lower than actual.

Status: Ongoing
Product Quantity: 19,429 systems
Code Information:

UDI-DI: 04049098048989 (M300), 04049098095778 (M300+); All Systems Running Software versions: VG2.4 and lower

Distribution Pattern:

Worldwide - US Nationwide distribution.

Voluntary or Mandated:

Voluntary: Firm initiated