Dynavox Systems Llc: Medical Device Recall in 2018 - (Recall #: Z-1304-2018)
See the recall detail below. You can also see other recalls from the same firm in 2018.
Dynavox T10 Speech Generating Device ( including Batteries)
Class II
The battery is integral to the device. If the device sits unused and is not charged for a lengthy period of time (a few months) the battery will go into a deep discharge state and the battery will expire. The battery may out-gas which may be described as swelled or puffed, as gas is trapped inside the pack surrounding the cells. This battery may cause the device case to separate a small amount.
TS1308 TS1411 Part Numbers: T10: 750394 Date Codes are part of the device s unique Serial Number and represent the year and month the device was manufactured.
Nationally
Voluntary: Firm initiated