Ecolab Inc: Medical Device Recall in 2016 - (Recall #: Z-2744-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

Microtek Franz Angiography Drape with Radiation Shield, Item # RBA202, and Microtek Angiography Drape with Radiation Shield and Pouches, Item # RBA203. Sterile EO. Microtek Medical, Inc., Columbus, MS 39702.

Product Classification:

Class II

Date Initiated: July 14, 2016
Date Posted: September 14, 2016
Recall Number: Z-2744-2016
Event ID: 74710
Reason for Recall:

Ecolab is recalling 2 models of Angiography Drapes with Radiation Shields because the sterile packaging may contain small holes in the film of the pouch.

Status: Terminated
Product Quantity: 6750 eaches (675 cases)
Code Information:

RBA202 Lot numbers: D111574, D111654, D111724, D111864, D112024, D112094, D112224, D120164, D120454, D120544, D121144, D121314, D121674, D121944, D122204, D122554, D122854, D123324, D130354, D130454, D130914, D131704, D131994, D132464, D132974, D140834, D141204, D141774, D142244, D150404, D151334, D151464, D152304, D152364, D152454, D153084, D153204, D153494, D160544. RBA203 Lot numbers: D122194, D122354

Distribution Pattern:

US: OH, NY, CA, MN.

Voluntary or Mandated:

Voluntary: Firm initiated