Edward Lifesciences, Llc: Medical Device Recall in 2015 - (Recall #: Z-2776-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2015.

Data Source: FDA.
Product Description:

Edwards Pressure Monitoring Kit with TruWave disposable pressure transducer. A sterile, single-use kit that monitors pressures when attached to pressure monitoring catheters. For use on patients requiring intravascular, intracranial, or intrauterine pressure monitoring.

Product Classification:

Class II

Date Initiated: July 31, 2015
Date Posted: September 23, 2015
Recall Number: Z-2776-2015
Event ID: 71971
Reason for Recall:

Flow restriction issues in a trifurcated IV set included in the pressure transducer kits. A restriction of the flow of IV fluid would result in either an inability to prime the system before use or difficulty in flushing the line after a blood draw.

Status: Terminated
Product Quantity: 13,077 (U.S.) and 12,306 (International)
Code Information:

Model # Lot # Expiration Date PX2X3 59987950 3/12/2017 PX2X3 60008882 3/24/2017 PX2X3 60043561 5/1/2017 PX3X3272 59994227 3/8/2017 PX3X3272 60025936 4/15/2017 PX3X3AN 60030889 4/17/2017 PX3X3AN 60008883 3/26/2017 PXAK2066 60025499 2/27/2017 PXAVMP3 60036757 4/27/2017 PXMK1970 60001398 4/16/2017 PXMK2043 59994210 3/9/2017 PXMK2043 60001412 3/17/2017 PXMK2043 60020166 4/5/2017 PXMK2043 60030868 4/26/2017 PXMK2144 60019038 4/17/2017 PXMK2217 59994267 3/12/2017 PXMK2217 60030969 4/20/2017 PXMK2217 60020235 4/15/2017 PXVK0458 59983333 4/6/2017 PXVK1079 59994195 3/11/2017 PXVK1079 60020153 3/31/2017 PXVMP2X31 59998630 3/12/2017 PXVMP2X31 59999004 3/13/2017 PXVMP2X31 59999006 3/13/2017 PXVMP3X31 59994223 3/13/2017 PXVMP3X31 60049581 4/23/2017 PXVMP3X32 60025874 3/31/2017 T001746A 60036771 4/30/2017 T001759A 59976102 3/10/2017 T001759A 60014134 3/31/2017 T001759A 60020177 4/13/2017 T001759A 60049570 5/4/2017 T330V11A 60014160 3/29/2017 T333F00A 59987991 3/9/2017 T333F00A 60001468 3/16/2017 T333F00A 60022033 3/13/2017 T334041A 60008925 3/25/2017 T334041A 60049612 5/5/2017 T336547B 60008919 3/27/2017 T337305B 59987992 3/11/2017 T337305B 60001473 3/17/2017 T391314A 60020222 3/29/2017 T391T01A 60014163 3/30/2017 T395606B 60022034 3/13/2017 T395606B 60025971 4/14/2017 T430140A 60014177 3/31/2017 T430609A 60001486 3/16/2017 T430609A 60043616 4/27/2017 T431305A 60008942 3/25/2017 T431305A 60059909 5/8/2017 T440B20B 59994257 3/13/2017 T440B20B 60030953 4/17/2017 T440B21A 60025987 4/15/2017 T440B21A 59976182 3/11/2017 T440B26B 60030938 4/17/2017 T440B26B 60049616 5/4/2017 T443422A 60036750 4/29/2017 T443422A 60020223 4/15/2017 T449103B 60008929 3/24/2017 T449103B 60043599 4/28/2017 T450544A 60030944 4/20/2017 T494C00B 60020168 3/19/2017 T494C00B 60001415 3/25/2017 T494C00B 60043543 4/21/2017 T530116B 60022035 3/13/2017 T530116B 60008937 3/24/2017

Distribution Pattern:

Worldwide Distribution - US nationwide, Australia, Austria, Canada, China, Colombia, Croatia, Denmark, France, Germany, Great Britain, Italy, Netherlands, New Zealand, Sweden, and Slovenia.

Voluntary or Mandated:

Voluntary: Firm initiated