EKOS Corporation: Medical Device Recall in 2013 - (Recall #: Z-0133-2014)
See the recall detail below. You can also see other recalls from the same firm in 2013.
The EkoSonic Endovascular System employs high frequency (2-2.5 MHz), low power (0.585 watt average power per transducer) ultrasound to facilitate the delivery of physician-specified fluids in the peripheral vasculature and to deliver solutions into the pulmonary arteries. The EkoSonic Endovascular System is intended for the controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature and the infusion of solutions into the pulmonary arteries. The reusable EKOS EkoSonic Control Unit provides power to the system and the user interface for operator control. A reusable, non sterile Connector Interface Cable connects the EKOS EkoSonic Control Unit to the EkoSonic Endovascular Device.
Class II
The EkoSonic Control Unit is recalled because it had the incorrect power entry module installed. It is missing an AC filter to reduce the noise and electrical interference.
PT3B-783. PT3B represents the model number. Adding 783 creates the serial number.
Worldwide Distribution in Germany only.
Voluntary: Firm initiated