Elekta, Inc.: Medical Device Recall in 2019 - (Recall #: Z-0821-2019)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2019.

Data Source: FDA.
Product Description:

MOSAIQ¿ Oncology Information System and Sequencer MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used wherever radiotherapy and/or chemotherapy are prescribed.

Product Classification:

Class II

Date Initiated: December 13, 2018
Date Posted: February 20, 2019
Recall Number: Z-0821-2019
Event ID: 81785
Reason for Recall:

There is a potential that Wedge IDs were not included in the DICOM RT PLAN sent from the TPS, which will cause the field in MOSAIQ to be created with no wedge. Treatment of a field without the planned wedge will receive more dose than the treatment plan indicates.

Status: Ongoing
Product Quantity: 3,950 units total (1,747 US)
Code Information:

MOSAIQ Radiation Oncology version 1.0 and higher. Multi-ACCESS version 6.1 and higher. UDI 858164002084 858164002091 858164002107 858164002220 858164002237

Distribution Pattern:

U.S. Nationwide distribution to the following states AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the District of Columbia and Puerto Rico.

Voluntary or Mandated:

Voluntary: Firm initiated