Elekta, Inc.: Medical Device Recall in 2025 - (Recall #: Z-1162-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

MOSAIQ Oncology Information System

Product Classification:

Class II

Date Initiated: January 15, 2025
Date Posted: February 19, 2025
Recall Number: Z-1162-2025
Event ID: 96153
Reason for Recall:

Image offsets calculated by oncology Information System software may be incorrect for Linear Accelerator (LINACs) not characterized at IEC61217 scale.

Status: Ongoing
Product Quantity: 64 units
Code Information:

Software versions /UDI: 3.1.2.0 (01)07340201500026 3.1.3.0 (01)07340201500026; 3.1.3.1 (01)07340201500026; 3.1.3.2 (01)07340201500026; 3.1.3.3 (01)07340201500026; 3.1.3.4 (01)07340201500026; 3.2.1.0 (01)07340201500071; 3.2.1.1 (01)07340201500071; 3.2.1.2 (01)07340201500071; 3.2.1.3 (01)07340201500071; 3.2.1.4 (01)07340201500071

Distribution Pattern:

US: PA, AL, FL, IA UT OUS: Austria Belgium Bolivia Bosnia and Herzegowina Brazil Canada France Germany Hong Kong Italy Japan Korea, Republic of Malta Netherlands Panama South Africa Spain Taiwan Thailand Turkey Venezuela

Voluntary or Mandated:

Voluntary: Firm initiated