Elekta Instrument AB: Medical Device Recall in 2023 - (Recall #: Z-1309-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

Instructions for Use for the following Leksell Stereotactic System and Leksell Neurosurgical Instruments: (1) Part #A2800-26, Backlund Catheter Insertion Needle Kit III; (2) Part #A2800-15, Backlund Catheter Insertion Needle Kit I; (3) Part #A2600-01, Backlund Haematoma Evacuator Kit; (4) Part #A2200-01, Salcman Twist Drill Kit I; (5) Part #907801, Insertion Cannula Kit; (6) Part #50398-01, Catheter Inserter Needle 1,5; (7) Part #307165, Insertion Cannula 190 MM; (8) Part #60377-02, Twist Drill 3.2 MM; (9) Part #60377-01, Twist Drill 2.1 MM; (10) Part #50376-01, Salcman Twist Drill 2.1/3.2; (11) Part #14001050, Catheter Inserter Needle 2,6; and (12) Part #1002248, Frame G w Straight/Curved Front.

Product Classification:

Class II

Date Initiated: March 1, 2023
Date Posted: April 5, 2023
Recall Number: Z-1309-2023
Event ID: 91780
Reason for Recall:

The incorrect IFU was distributed with the devices.

Status: Ongoing
Product Quantity: 70 devices
Code Information:

UDI-DI: (1) 07340048301305; (2) 07340048300285; (3) 07340048301220; (4) 07340048301145; (5) 07340048300315; (6) 07340048308618; (7) 07340048308632; (8) 07340048308571; (9) 07340048308564; (10 07340048308601; (11) 07340048308625; and (12) 07340048306348.

Distribution Pattern:

US Nationwide distribution in the states of AZ, CA, CO, FL, GA, ID, IL, KS, MA, MD, MI, MN, MO, MS, NE, NH, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, WA, WI, and WV. There was also government distribution.

Voluntary or Mandated:

Voluntary: Firm initiated