Encore Medical, Lp: Medical Device Recall in 2017 - (Recall #: Z-1413-2017)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2017.
Data Source: FDA.
Product Description:
RSP Impaction Fixture
Product Classification:
Class II
Date Initiated: February 9, 2017
Date Posted: March 15, 2017
Recall Number: Z-1413-2017
Event ID: 76456
Reason for Recall:
During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.
Status: Terminated
Product Quantity: 626 units
Code Information:
109931L01, 109931L02, 115670L15, 115670L16, 128092L08, 128092L09, 137917L16, 167829L06, 52258L01, 52258L01A, 67428L01, 76386L01, 76386L02, 81722L01, 81722L02
Distribution Pattern:
US, South Korea, Australia, Canada, United Kingdom/Ireland, Germany
Voluntary or Mandated:
Voluntary: Firm initiated