Encore Medical, Lp: Medical Device Recall in 2018 - (Recall #: Z-0329-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

EMPOWR Porous Knee System TIBIAL PUNCH, LARGE, REF 801-05-204, NON STERILE, Qty 1 For use in orthopedic surgery.

Product Classification:

Class II

Date Initiated: December 12, 2017
Date Posted: January 17, 2018
Recall Number: Z-0329-2018
Event ID: 78727
Reason for Recall:

The Locking Punch Guide was not sufficiently guiding the large press fit punch, causing the implant to be placed 2-3mm more posterior than intended.

Status: Terminated
Product Quantity: 2 units
Code Information:

Lot 252564

Distribution Pattern:

One medical device distributor in California.

Voluntary or Mandated:

Voluntary: Firm initiated