Encore Medical, LP: Medical Device Recall in 2020 - (Recall #: Z-3022-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
RSP Standard HUMERAL SOCKET INSERT, 40 MM plus 4MM, HXe-Plus, Part No. 509-00-440
Class II
Product may be mislabeled with the size etched onto products differing from the size information shown on patient sticker labels. Risks associated with this issue may include the Surgeon implanting the incorrect size; significant physical impairment, discomfort and/or disruption of treatment of underlying condition; and the patient experiencing dislocation resulting in revision surgery.
Part 509-00-440, Lot Numbers 387P1042A and 387P1045A, UDI: 00888912144599
Distributed in the U.S. to: VA, MN, IN, RI, PA, NY, WA, KS, OK, FL, GA and CA
Voluntary: Firm initiated