Encore Medical, LP: Medical Device Recall in 2023 - (Recall #: Z-1187-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

EMPOWR VVC KNEE TIBIAL INSERT, SIZE 6, 22MM, REF: 346-22-706; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 7, 22MM, REF: 346-22-707; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 8, 22MM, REF: 346-22-708; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 9, 22MM, REF: 346-22-709; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 10, 22MM, REF: 346-22-710; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 11, 22MM, REF: 346-22-711; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 5 BRIDGE UP, 22MM, REF: 346-22-755; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 6 BRIDGE DOWN, 22MM, REF: 346-22-766; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 2, 25MM, REF: 346-25-702; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 3, 25MM, REF: 346-25-703; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 4, 25MM, REF: 346-25-704; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 5, 25MM, REF: 346-25-705; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 5 BRIDGE UP, 25MM, REF: 346-25-755; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 6 BRIDGE DOWN, 25MM, REF: 346-25-766

Product Classification:

Class II

Date Initiated: December 22, 2022
Date Posted: March 8, 2023
Recall Number: Z-1187-2023
Event ID: 91494
Reason for Recall:

Tibial insert implant large boxes were labeled small and the small boxes were labeled large. If the incorrect labeling is identified this could lead to surgical delay and if the surgeon does not recognize the size discrepancy between the insert label and the pin label, the surgeon may implant the incorrect pin size.

Status: Ongoing
Product Quantity: 216
Code Information:

REF/UDI-DI/Lot: 346-22-706/ 00190446219378/484V1013, 484V1014, 484V1015, 484V1016, 484V1017, 484V1018; 346-22-707/ 00190446219385/485V1001, 485V1002, 485V1003, 485V1004; 346-22-708/ 00190446219392/486V1005, 486V1006, 486V1007, 486V1008; 346-22-709/ 00190446219408/487V1001, 487V1002, 487V1003, 487V1004; 346-22-710/ 00190446219415/488V1003, 488V1004, 488V1005, 488V1006; 346-22-711/ 00190446219422/489V1003, 489V1004, 489V1005, 489V1006; 346-22-755/ 00190446219439/490V1001, 490V1002, 490V1003, 490V1004; 346-22-766/ 00190446219477/491V1002, 491V1003, 491V1004; 346-25-702/ 00190446219507/492V1002, 492V1003, 492V1004, 492V1005; 346-25-703/ 00190446219569/493V1001, 493V1002, 493V1003, 493V1004; 346-25-704/ 00190446219583/494V1001, 494V1002, 494V1003, 494V1004; 346-25-705/ 00190446219590/495V1001, 495V1002, 495V1003, 495V1004; 346-25-755/ 00190446219781/ 502V1001, 502V1002, 502V1003; 346-25-766/ 00190446219798/503V1003, 503V1004, 503V1005, 503V1006, 503V1007;

Distribution Pattern:

US: MN, IL, IN, NY, FL

Voluntary or Mandated:

Voluntary: Firm initiated