Encore Medical, LP: Medical Device Recall in 2023 - (Recall #: Z-2606-2023)
See the recall detail below. You can also see other recalls from the same firm in 2023.
DJO Surgical: Reverse Shoulder System (RSP) Semi Humeral Socket Insert, 32mm, REF: 508-01-032; Baseplate, Glenoid HA-Coat, RSP, 6.5mm X 30mm, REF: 508-32-104; Reverse Shoulder System (RSP) Glenoid, Neutral Head W/ Retaining Screw, 36mm, REF: 508-36-101
Class II
Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to revision surgery
REF/UDI-DI/Lot/Expiration: 508-01-032/00888912024693/859C1190/April 29, 2027; 508-32-104/00888912024754/866C3153/May 25, 2029 and 866C3147/April 3, 2029; 508-36-101/869C3539/April 30, 2028
US: AL, AZ, CA, CO, FL, IL, IN, KS, KY, LA, MA, MI, MN, MO, NC, NJ, NY, OH, PA, PR, RI, SC, TX, UT, VA, WA. OUS: Sweden, Australia, France, UK, Colombia, Germany
Voluntary: Firm initiated