ESAOTE S.P.A: Medical Device Recall in 2020 - (Recall #: Z-3052-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
AC/DC POWER SUPPLY for 7400, Part Number 151004300, XP Power Model Number AHM180PS19-XE0338, AC/DC Adapter associated with the following portable ultrasound diagnostic systems: The AC/DC adapter is intended for use with the following portable ultrasound systems: MyLabAlpha (101740000) MyLabGamma (101741000) MyLabOmega (101740004) MyLabSigma (101741004) MyLabSigma elite (101741007) MyLabAlphaVET (101740001) MyLabGammaVET (101741001) MyLabDelta (100743000) MyLabDeltaVET (100743001) MyLabOmegaVET (101740005) MyLabSigmaVET (101741005)
Class II
The AC/DC adapter (power supply) of the portable ultrasound systems listed below that may result in the connector heating up, generating smoke and melting. This condition is caused by the mechanical stress of the connection between the AC/DC power adapter and the mains cable over the time of use.
Serial Numbers: Below K17510001
US Nationwide
Voluntary: Firm initiated