Estill Medical Technologies, Inc: Medical Device Recall in 2014 - (Recall #: Z-0101-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

Ultra Battery 1 Model: TA-UB1 NSN: 6515-01-592-5803 The Ultra Battery 1 (TA-UB1) is a rechargeable power source for the Thermal Angel Blood and IV Fluid Infusion Warmer. The TA-UB1 battery was designed to provide sufficient power for the Thermal Angel to warm and deliver one liter of blood or IV fluid infusion to a trauma patient.

Product Classification:

Class II

Date Initiated: July 25, 2014
Date Posted: October 29, 2014
Recall Number: Z-0101-2015
Event ID: 69242
Reason for Recall:

Estill has discovered that a fully discharged Ultra Battery 1 (TA-UB1) will reach 96.9% of its capacity during a 60 minute charging period. If you have a fully discharged Ultra Battery 1 (TA-UB1), in order to achieve a 100% of charge capacity you should charge it for a period of 90 minutes. In addition, the charging voltage range of 10-36V specified on the Ultra Battery 1 (TA-UB1) label and in t

Status: Terminated
Product Quantity: 6,886
Code Information:

06-2011; 07-2011; 08-2011; 09-2011; 03-2012; 04-2012; 05-2012; 06-2012; 07-2012; 08-2012; 01-2013; 03-2013; 08-2013; 09-2013; 10-2013; 11-2013; 12-2013; 01-2014; 02-2014; 04-2014; 07-2012; 02-2014

Distribution Pattern:

Worldwide Distribution - USA (nationwide) and Internationally to Serbia.

Voluntary or Mandated:

Voluntary: Firm initiated