Ethicon Endo-Surgery Inc: Medical Device Recall in 2018 - (Recall #: Z-1886-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

5mm Endopath Xcel with Optiview Technology, Bladeless Trocar with Stability Sleeve, 5mm, 75mm shaft length, 2B5ST

Product Classification:

Class II

Date Initiated: March 26, 2018
Date Posted: May 23, 2018
Recall Number: Z-1886-2018
Event ID: 79801
Reason for Recall:

Use of the product with insufficient lubrication may prolong surgery due to trocar replacement. In addition, insufficient lubrication could result in a remote chance of introducing a minute particle if repeated insertions of the obturator or laparoscopic device damage the seal.

Status: Terminated
Product Quantity: 14873
Code Information:

Lots P4T745 P4T88Y P4T906 P4T91H P4TA60 P4TC8M P4TD0G P4TD2K R40015 R4011E R4025E R4035E R4042Z R4047X R4065E

Distribution Pattern:

Nationwide domestic distribution, including PR. International distribution to Saudi Arabia, United Arab Emirates, Belgium, Singapore, Taiwan, South Korea, Canada, Chile, Mexico, Japan, and Argentina.

Voluntary or Mandated:

Voluntary: Firm initiated