Ethicon Endo-Surgery Inc: Medical Device Recall in 2021 - (Recall #: Z-2355-2021)
See the recall detail below. You can also see other recalls from the same firm in 2021.
HARMONIC HD 1000i Shears (20cm Shaft Length)-indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. Product Code: HARHD20
Class II
An internal component may be cracked and become lodged behind the energy button potentially resulting in continuous activation of the device, may result in burns
Lot Codes: U95126 U9571N U9526Z U95A80 U9543P U95T0X U9550G GTIN: 10705036015048
Worldwide distribution - US Nationwide distribution and the countries of Austria, Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czechia, Denmark, Estonia, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, United Arab Emirates.
Voluntary: Firm initiated